4.7 Article

Voltammetric aptasensor for sulfadimethoxine using a nanohybrid composed of multifunctional fullerene, reduced graphene oxide and Pt@Au nanoparticles, and based on direct electron transfer to the active site of glucose oxidase

期刊

MICROCHIMICA ACTA
卷 186, 期 1, 页码 -

出版社

SPRINGER WIEN
DOI: 10.1007/s00604-018-3127-5

关键词

Poly(diallyldimethylammonium chloride); Aptamer; Electrochemical detection; Glucose oxidase

资金

  1. National Natural Science Foundation of China [81601856]
  2. Third Batch of Young Backbone Teachers Funding Program in Colleges and Universities of Chongqing City [53]
  3. Funds for Young Science and Technology Talent Cultivation Plan of Chongqing City [cstc2014kjrc-qnrc00004]

向作者/读者索取更多资源

This work describes a voltammetric and ultrasensitive aptasensor for sulfadimethoxine (SDM). It is based on signal amplification by making use of a multifunctional fullerene-doped reduced graphene oxide nanohybrid. The nanohybrid was coated with poly(diallyldimethylammonium chloride) to obtain a material (P-C-60-rGO) with large specific surface area and a unique adsorption ability for loading it with glucose oxidase (GOx). The coating also facilitates the direct electron transfer between the active site of GOx and the glassy carbon electrode (GCE). The P-C-60-rGO were then modified with Pt@Au nanoparticles, and the thiolated SDM-binding aptamer was immobilized on the nanoparticles. On exposure of the modified GCE to a solution containing SDM, it binds to the aptamer. The results were recorded through the signal responses generated from the redox center of GOx (FAD/FADH(2)) by cyclic voltammetry at a scan rate of 100mVs(-1) from -0.25 to -0.65V. Accordingly, The sensor has good specificity and stability, and response is linearin the 10fgmL(-1) to 50ngmL(-1) SDM concentration range with a detection limit of 8.7fgmL(-1).

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据