Technical Note: Intrinsic raw data-based CT misalignment correction without redundant data

期刊

MEDICAL PHYSICS
卷 46, 期 1, 页码 173-179

出版社

WILEY
DOI: 10.1002/mp.13254

关键词

-

资金

  1. Deutsche Forschungsgemeinschaft (DFG) [SA 2776/1-1]
  2. NVIDIA Corporation

向作者/读者索取更多资源

Purpose CT image reconstruction requires accurate knowledge of the used geometry or image quality might be degraded by misalignment artifacts. To overcome this issue, an intrinsic method, that is, a method not requiring a dedicated calibration phantom, to perform a raw data-based misalignment correction for CT is proposed herein that does not require redundant data and hence is applicable to measurements with less than 180 circle plus fan-angle of data. Methods The forward projection of a volume reconstructed from a misaligned geometry resembles the acquired raw data if no redundant data are used, that is, if less than 180 circle plus fan-angle are used for image reconstruction. Hence, geometric parameters cannot be deduced from such data by an optimization of the geometry-dependent raw data fidelity. We propose to use a nonlinear transform applied to the reconstructed volume to introduce inconsistencies in the raw data that can be employed to estimate geometric parameters using less than 180 circle plus fan-angle of data. The proposed method is evaluated using simulations of the FORBILD head phantom and using actual measurements of a contrast-enhanced scan of a mouse acquired using a micro-CT. Results Noisy simulations and actual measurements demonstrate that the proposed method is capable of correcting for artifacts arising from a misaligned geometry without redundant data while ensuring raw data fidelity. Conclusions The proposed method extends intrinsic raw data-based misalignment correction methods to an angular range of 180 circle or less and is thus applicable to systems with a limited scan range.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据