Technical Note: U-net-generated synthetic CT images for magnetic resonance imaging-only prostate intensity-modulated radiation therapy treatment planning

期刊

MEDICAL PHYSICS
卷 45, 期 12, 页码 5659-5665

出版社

WILEY
DOI: 10.1002/mp.13247

关键词

deep learning; IMRT; prostate; synthetic CT; U-net

资金

  1. Elekta RD grant

向作者/读者索取更多资源

Purpose Clinical implementation of magnetic resonance imaging (MRI)-only radiotherapy requires a method to derive synthetic CT image (S-CT) for dose calculation. This study investigated the feasibility of building a deep convolutional neural network for MRI-based S-CT generation and evaluated the dosimetric accuracy on prostate IMRT planning. Methods A paired CT and T2-weighted MR images were acquired from each of 51 prostate cancer patients. Fifteen pairs were randomly chosen as tested set and the remaining 36 pairs as training set. The training subjects were augmented by applying artificial deformations and feed to a two-dimensional U-net which contains 23 convolutional layers and 25.29 million trainable parameters. The U-net represents a nonlinear function with input an MR slice and output the corresponding S-CT slice. The mean absolute error (MAE) of Hounsfield unit (HU) between the true CT and S-CT images was used to evaluate the HU estimation accuracy. IMRT plans with dose 79.2 Gy prescribed to the PTV were applied using the true CT images. The true CT images then were replaced by the S-CT images and the dose matrices were recalculated on the same plan and compared to the one obtained from the true CT using gamma index analysis and absolute point dose discrepancy. Results The U-net was trained from scratch in 58.67 h using a GP100-GPU. The computation time for generating a new S-CT volume image was 3.84-7.65 s. Within body, the (mean +/- SD) of MAE was (29.96 +/- 4.87) HU. The 1%/1 mm and 2%/2 mm gamma pass rates were over 98.03% and 99.36% respectively. The DVH parameters discrepancy was less than 0.87% and the maximum point dose discrepancy within PTV was less than 1.01% respect to the prescription. Conclusion The U-net can generate S-CT images from conventional MR image within seconds with high dosimetric accuracy for prostate IMRT plan.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据