Accuracy of 3D surface scanners for clinical torso and spinal deformity assessment

期刊

MEDICAL ENGINEERING & PHYSICS
卷 63, 期 -, 页码 63-71

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.medengphy.2018.11.004

关键词

3D Scanning; Scoliosis; Torso Deformity; Surface Deviations; Accuracy; Reproducibility of results; Surface topography

资金

  1. Queensland University of Technology
  2. Synthes
  3. Mater Foundation
  4. Medtronic

向作者/读者索取更多资源

Externally visible deformities are cosmetic features of great concern for Adolescent Idiopathic Scoliosis (AIS) patients. Current assessment techniques for AIS do not fully encompass the external deformity. A non-invasive method capable of capturing superficial anatomy, such as 3D scanning, would enable better qualitative and quantitative evaluation of cosmesis. This study aimed to quantify the accuracy of commonly available scanners, in assessing posterior asymmetry in AIS. The technique of 3D surface deviation analysis was proposed as a suitable method for comparing the models created by each scanner. Eight plaster cast moulds manufactured to create braces for AIS patients were used as test samples. Four 3D scanners were selected: Solutionix RexScan CS+; Artec Eva; Microsoft Kinect V1; iPhone with 123D Catch App. These scanners were selected from those available as representative of a range of scanning technologies. Each cast was scanned and 3D models created. A simulated rib hump measurement was obtained and the surface-to-surface deviations between the Solutionix scan and all other scans were determined. The Solutionix scanner is a metrology scanner of very high quality and so it was selected as the reference. Surface-to-surface deviations were calculated in the positive and negative directions separately to specifically identify size and volume inaccuracies created by the scans. Surface deviations showed excellent agreement between the Solutionix and the Eva with deviations of +0.17 +/- 0.17 mm (Eva regions larger) and 0.20 +/- 0.32 mm (Eva regions smaller) (mean SD). The Kinect showed lower agreement (+1.58 +/- 1.50 mm and 0.58 +/- 0.58 mm). The iPhone scans were not able to be scaled to the correct size, so were excluded. Rib hump measurements with all scanners were within clinical measurement variability (+/- 4.9 deg) of the known values. These commercially available 3D scanners are capable of imaging torso shape in 3D and deriving clinically relevant external deformity measures. The non-invasive 3D topographic information provided can be used to improve assessment of torso shape in spinal deformity patients. (C) 2018 IPEM. Published by Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据