期刊
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES
卷 1104, 期 -, 页码 134-140出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.jchromb.2018.11.017
关键词
Avermectins; Milbemycines; LC-MS/MS; Fish; QuEChERS
In this study, the simultaneous determination of avermectins (emamectin, eprinomectin, abamectin, doramectin and ivermectin) and milbemycines (moxidectin) in fish tissue with LC-ESI-MS/MS, was studied. Optimum chromatographic separation of target analytes was achieved using a Waters Acquity UPLC BEH C-18 (100 mm x 2.1 mm, 1.7 mm) analytical column, operated at 40 degrees C and the composition of the mobile phase used was (A): ACN-MeOH (0.1% HCOOH) (1:1) and (B): 1 mM HCOONH4 (0.1% HCOOH). Various mobile phases were tested and the effect of the mobile phase composition on the analytes ionization was thoroughly examined in an extensive ionization study, aiming to increase the analytes' sensitivity. Deuterated ivermectin (IVR-d(2)) was used as an internal standard (IS). Avermectins and milbemycine's extraction from the fish matrix was conducted with acidified ACN (0.1% HCOOH), followed by QuEChERS methodology. The method developed herein was validated according to the European Legislation requirements (Commission Decision 657/2002/EC) and recoveries ranged from 86 to 106% for all target analytes, with relative standard deviations < 20%. LODs ranged from 0.07 mu g/kg (emamectin) to 1.3 mu g/kg (doramectin), indicating the excellent sensitivity of the method. The developed methodology was successfully applied to fish samples obtained from aquaculture.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐