4.3 Article

Precise Cobb Angle Measurement System Based on Spinal Images Merging Function

期刊

IRBM
卷 39, 期 5, 页码 343-352

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.irbm.2018.09.002

关键词

-

资金

  1. University of Ulsan

向作者/读者索取更多资源

Background: Convenient and precise measurement of the Cobb angle using a small size X-ray detector has been required for local clinics. Methods: Cobb angle measurement system using a conventional X-ray source and detector is proposed for accurate Cobb angle measurement. The system consists of a conventional X-ray source, a ruler-added Xray table, a conventional X-ray detector, and an image processing program. The X-ray table has the lead ruler patterns. The patterns remain white ruler patterns on X-ray images. The proposed image processing program merges the three spinal X-ray images into one whole spinal X-ray image by detecting the ruler patterns on the three spinal X-ray images. Results: In order to evaluate our program. Cobb angle measured in the merged image is compared with Cobb angle measured in the X-ray image taken by a large X-ray detector. Average of difference between them is 2.251 degree and standard deviation is 1.339. Conclusion: The developed measurement system demonstrated its measurement performance accurately and practically. (C) 2018 AGBM. Published by Elsevier Masson SAS. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据