Is calibration the cause of variation in liquid chromatography tandem mass spectrometry testosterone measurement?

期刊

ANNALS OF CLINICAL BIOCHEMISTRY
卷 50, 期 4, 页码 368-370

出版社

SAGE PUBLICATIONS INC
DOI: 10.1177/0004563212474563

关键词

Mass spectrometry; laboratory methods; steroid hormones; analytes

向作者/读者索取更多资源

Background: Testosterone measurement by liquid chromatography tandem mass spectrometry (LC-MS/MS) is well accepted as the preferred technique for the analysis of testosterone. Variation is seen between assays and this may be due to differences in calibration as commercial calibrators for this assay are not readily available. We investigated the effects calibration in routine clinical LC-MS/MS assays. Methods: All LC-MS/MS users that were registered with the UKNEQAS external quality assurance scheme for testosterone were invited to take part in the study. A set of seven serum samples and serum-based calibrators were sent to all laboratories that expressed an interest. The laboratories were instructed to analyse all samples using there own calibrators and return the results and a method questionnaire for analysis. Results: Fifteen laboratories took part in the study. There was no consensus on supplier of testosterone or matrix for the preparation of calibrators and all were prepared in-house. Also, a wide variety of mass spectrometers, internal standards, chromatography conditions and sample extractions were used. The variation in results did not improve when the results were corrected with a common calibrator. Conclusions: The variation in results obtained could not be attributed to variations in calibrators. The differences in methodologies between laboratories must be the reason for this variation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据