4.3 Article

A microfabricated fixed path length silicon sample holder improves background subtraction for cryoSAXS

期刊

JOURNAL OF APPLIED CRYSTALLOGRAPHY
卷 48, 期 -, 页码 227-237

出版社

INT UNION CRYSTALLOGRAPHY
DOI: 10.1107/S1600576714027782

关键词

cryoSAXS; biological small-angle X-ray scattering; sample holders; 100 K; solution scattering; fixed path length

资金

  1. National Science Foundation [DBI-1152348, ECCS-0335765, DGE-1144153]
  2. National Institutes of Health/National Institute of General Medical Sciences under NSF [DMR-0936384]
  3. National Institute of General Medical Sciences, National Institutes of Health [GM-103485]
  4. Div Of Biological Infrastructure
  5. Direct For Biological Sciences [1152348] Funding Source: National Science Foundation

向作者/读者索取更多资源

The application of small-angle X-ray scattering (SAXS) for high-throughput characterization of biological macromolecules in solution is limited by radiation damage. By cryocooling samples, radiation damage and required sample volumes can be reduced by orders of magnitude. However, the challenges of reproducibly creating the identically sized vitrified samples necessary for conventional background subtraction limit the widespread adoption of this method. Fixed path length silicon sample holders for cryoSAXS have been microfabricated to address these challenges. They have low background scattering and X-ray absorption, require only 640nl of sample, and allow reproducible sample cooling. Data collected in the sample holders from a nominal illuminated sample volume of 2.5nl are reproducible down to q similar or equal to 0.02 angstrom(-1), agree with previous cryoSAXS work and are of sufficient quality for reconstructions that match measured crystal structures. These sample holders thus allow faster, more routine cryoSAXS data collection. Additional development is required to reduce sample fracturing and improve data quality at low q.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据