Accuracy of urine pH testing in a regional metabolic renal clinic: is the dipstick accurate enough?

期刊

UROLITHIASIS
卷 41, 期 2, 页码 129-132

出版社

SPRINGER
DOI: 10.1007/s00240-013-0546-y

关键词

Urine; pH; Nephrolithiasis; Clinical management

资金

  1. National Institute for Health Research Cambridge Biomedical Research Centre
  2. Welcome Trust [088489]

向作者/读者索取更多资源

Urine pH is a useful marker for assessing treatment need and efficacy in patients with nephrolithiasis. Though the gold standard of measurement is with a pH electrode, dipsticks offer the convenience of cost, ease of use, and the possibility of patients measuring their own values outside the clinic. The aim of this study was to determine whether dipsticks offer the same accuracy as the electrode. Paired measurements of freshly voided urine pH with both electrode and dipstick were analysed in a multidisciplinary renal clinic. We found that although there was a high Pearson correlation between the samples (0.89, p = 0.001), urine dipstick measurements carried an approximately 1 in 4 risk of producing clinically significant differences (pH differences > 0.5 pH unit) from meter values. We also found that at high and low urine pH, the dipstick tended to over- and underestimate true pH readings, respectively. Examining the values in the 98 patients where a need for pharmacological urinary pH manipulation was indicated by the true pH, we found 14 who would not have been appropriately treated, and 5 who would have been unnecessarily medicated, if the stick pH value had been used. We conclude that dipstick pH measurement is insufficiently reliable for guiding clinical decision-making.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据