4.0 Article

Quality and Diagnostic Utility of Mydriatic Smartphone Photography: The Smartphone Ophthalmoscopy Reliability Trial

期刊

OPHTHALMIC SURGERY LASERS & IMAGING RETINA
卷 46, 期 6, 页码 631-637

出版社

SLACK INC
DOI: 10.3928/23258160-20150610-06

关键词

-

资金

  1. Wills Eye Hospital Innovations Grant

向作者/读者索取更多资源

BACKGROUND AND OBJECTIVE: Establish quality and diagnostic utility of mydriatic smartphone ophthalmoscopy (SO) fundus images compared to fundus camera (FC) images. PATIENTS AND METHODS: In this prospective, cross-sectional study, 94 consecutive patients in an urban eye emergency department underwent SO and FC fundus imaging via one of three study arms: medical student 1 (MS1), medical student 2 (MS2), and ophthalmology resident (OR). Images of 188 eyes were graded for overall quality by two masked reviewers, and observed critical fundus findings were compared to dilated fundus examination documentation. RESULTS: SO images were higher quality in the OR arm than in the MS1 and MS2 arms (P < .017). There were no differences in FC image quality between photographers (all P > .328). In the OR arm, SO images detected 74.3% of critical fundus findings, whereas FC images detected 77.1%. CONCLUSION: SO produces fundus images approaching the quality and diagnostic utility of traditional FC photographs.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据