4.4 Article

Left ventricular assist device malposition interrogated by 4-D cine computed tomography

期刊

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.jcct.2011.01.009

关键词

Cine CT; Left ventricular assist device

向作者/读者索取更多资源

67-year-old female with left ventricular assist device (LVAD) presented with recurrent low-flow alarms. No clear etiology could be determined by history, or evaluation with radiograph and echocardiogram. Computed tomographic (CT) imaging with 3-D and 4-D assessment identified the abnormality as steep angulation of the inflow cannula and partial obstruction by the adjacent anterior wall, likely in part caused by recovered left ventricular function and reverse remodeling. Improved left ventricle size and function was correlated by semi-automated analysis of CT data, which also indicated mild right ventricle dilation and systolic dysfunction. LVAD explantation was performed, and has been well tolerated by the patient. Echocardiography remains the primary imaging modality to assess patients post LVAD placement, but in this instance CT provided valuable information to identify the abnormality and help direct patient management. CT assessment in patients with LVAD additionally provides valuable information prior to redo sternotomy for pump explantation, revision, or transplantation. (C) 2011 Society of Cardiovascular Computed Tomography. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据