4.4 Article

Evaluation of a New Immunochromatographic Assay Kit for the Rapid Detection of Norovirus in Fecal Specimens

期刊

ANNALS OF LABORATORY MEDICINE
卷 32, 期 1, 页码 79-81

出版社

KOREAN SOC LABORATORY MEDICINE
DOI: 10.3343/alm.2012.32.1.79

关键词

Immunochromatographic assay; Norovirus; Sensitivity; Specificity

向作者/读者索取更多资源

Rapid and accurate detection of norovirus is essential for the prevention and control of norovirus outbreaks. This study compared the effectiveness of a new immunochromatographic assay kit (SD BIOLINE Norovirus; Standard Diagnostics, Korea) and real-time reverse transcription-PCR (RT-PCR) for detecting norovirus in fecal specimens. Compared with real-time RT-PCR, the new assay had sensitivity, specificity, positive predictive value, and negative predictive value of 76.5% (52/68), 99.7% (342/343), 98.1% (52/53), and 95.5% (342/358), respectively. The sensitivity of the assay was 81.8% (18/22) for GII. 3 and 75.7% (28/37) for GII. 4. None of the 38 enteric virus-positive specimens (3 for astrovirus, 5 for enteric adenovirus, and 30 for rotavirus) tested positive in the cross-reactivity test performed by using this assay. The new immunochromatographic assay may be a useful screening tool for the rapid detection of norovirus in sporadic and outbreak cases; however, negative results may require confirmatory assays of greater sensitivity.