4.7 Article

Novel amperometric glucose biosensor based on MXene nanocomposite

期刊

SCIENTIFIC REPORTS
卷 6, 期 -, 页码 -

出版社

NATURE PORTFOLIO
DOI: 10.1038/srep36422

关键词

-

资金

  1. King Abdullah University of Science and Technology
  2. Ramanujan Fellowship, Department of Science and Technology (DST), Govt. of India
  3. CSIR-NIIST Thiruvananthapuram, India

向作者/读者索取更多资源

A biosensor platform based on Au/MXene nanocomposite for sensitive enzymatic glucose detection is reported. The biosensor leverages the unique electrocatalytic properties and synergistic effects between Au nanoparticles and MXene sheets. An amperometric glucose biosensor is fabricated by the immobilization of glucose oxidase (GOx) enzyme on Nafion solubilized Au/MXene nanocomposite over glassy carbon electrode (GCE). The biomediated Au nanoparticles play a significant role in facilitating the electron exchange between the electroactive center of GOx and the electrode. The GOx/Au/MXene/Nafion/GCE biosensor electrode displayed a linear amperometric response in the glucose concentration range from 0.1 to 18 mM with a relatively high sensitivity of 4.2 mu AmM-1 cm(-2) and a detection limit of 5.9 mu M (S/N = 3). Furthermore, the biosensor exhibited excellent stability, reproducibility and repeatability. Therefore, the Au/MXene nanocomposite reported in this work is a potential candidate as an electrochemical transducer in electrochemical biosensors.