4.7 Article

Quantitative blood flow velocity imaging using laser speckle flowmetry

期刊

SCIENTIFIC REPORTS
卷 6, 期 -, 页码 -

出版社

NATURE PUBLISHING GROUP
DOI: 10.1038/srep25258

关键词

-

资金

  1. Microscan B.V. (Amsterdam)
  2. Gerbrand de Jong Fonds (the Netherlands)
  3. Dutch Technology Foundation STW (iMIT-Instruments for Minimally Invasive Techniques, project OBAMA)
  4. Dutch Technology Foundation STW (iMIT-Instruments for Minimally Invasive Techniques, project PROSPECT)

向作者/读者索取更多资源

Laser speckle flowmetry suffers from a debated quantification of the inverse relation between decorrelation time (tau(c)) and blood flow velocity (V), i.e. 1/tau(c) = alpha V. Using a modified microcirculation imager (integrated sidestream dark field - laser speckle contrast imaging [SDF-LSCI]), we experimentally investigate on the influence of the optical properties of scatterers on alpha in vitro and in vivo. We found a good agreement to theoretical predictions within certain limits for scatterer size and multiple scattering. We present a practical model-based scaling factor to correct for multiple scattering in microcirculatory vessels. Our results show that SDF-LSCI offers a quantitative measure of flow velocity in addition to vessel morphology, enabling the quantification of the clinically relevant blood flow, velocity and tissue perfusion.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据