The potential of an inexpensive dithizone immobilized on ammonium lauryl sulfate alumina adsorbent to separate and preconcentrate traces of mercury from vegetables and ground water has been evaluated. A cartridge filled with 100 mg of adsorbent was incorporated to an on-line vapor generation accessory system. Traces of mercury in sample solutions were pumped through the cartridges for 2 min (flow rate of 5 mL min(-1)) and then eluted from the cartridges with 1 M hydrochloric acid for 1 min at a flow rate of 1 mL min(-1). At the same time that mercury is eluted from the cartridges, the eluate is mixed with a 0.02% (m/v) sodium borohydride solution and the mercury cold vapor is swept through the atomization cell with a nitrogen flow of 50 mL min(-1). The overall procedure (mercury separation-preconcentration and mercury cold vapor generation-atomization) is completed in around 3 min. The developed method achieves an enrichment factor of 125 with a limit of detection of 0.12 ng mL(-1). The repeatability and the accuracy of the overall procedure were successfully assessed.