4.5 Article

Paclitaxel-coated balloon in combination with bare metal stent for treatment of de novo coronary lesions: an optical coherence tomography first-in-human randomised trial, balloon first vs. stent first

期刊

EUROINTERVENTION
卷 7, 期 6, 页码 711-722

出版社

EUROPA EDITION
DOI: 10.4244/EIJV7I6A114

关键词

coronary stenosis; angioplasty; drug-coat balloon; paclitaxel; optical coherence tomography

资金

  1. Lutonix Inc, Maple Grove, MA, USA

向作者/读者索取更多资源

Aims: To test the efficacy of sequential application of drug-coated balloon (DCB) and bare metal stent (BMS) for treatment of de novo coronary lesions, comparing the sequence of application (DCB first vs. BMS first). Methods and results: In a multicentre pilot trial, 26 patients with de 1101,0 coronary lesions were randomised to receive a paclitaxel-coated balloon application followed by BMS implantation (DCB first) or vice versa (BMS first). Quantitative coronary angiography (QCA) and optical coherence tomography (OCT) were performed post-procedure and at six months, with OCT % neointimal volume obstruction as primary endpoint. Longitudinal geographical miss was only observed in DCB first (23.1 vs. 0.0%, p=0.220). Implantation of BM S first resulted in fewer malapposed struts (p=0.013) but similar coverage at six months. No significant difference was found regarding the primary endpoint (25.5 vs. 24.9%, p=0.922), mean thickness of coverage (261 vs. 2251.1m, p=0.763), late loss (0.53 vs. 0.45 mm, p=0.833), binary restenosis (27.3 vs. 16.7% in-segment, p=0.640) or clinical endpoints. Conclusions: Sequential application of DCB and not pre-mounted BMS for treatment of de novo coronary lesions results in efficient inhibition of neointimal hyperplasia. The sequence of application (DCB first vs. BMS first) does not seem to influence the outcome, except for better apposition in BMS first.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据