4.6 Article

Simultaneous multiple-emitter fitting for single molecule super-resolution imaging

期刊

BIOMEDICAL OPTICS EXPRESS
卷 2, 期 5, 页码 1377-1393

出版社

OPTICAL SOC AMER
DOI: 10.1364/BOE.2.001377

关键词

-

资金

  1. NIH [R21RR024438, 1P50GM085273, 2U54RR022241]
  2. NSF [0954836]
  3. Division Of Physics
  4. Direct For Mathematical & Physical Scien [0954836] Funding Source: National Science Foundation

向作者/读者索取更多资源

Single molecule localization based super-resolution imaging techniques require repeated localization of many single emitters. We describe a method that uses the maximum likelihood estimator to localize multiple emitters simultaneously within a single, two-dimensional fitting sub-region, yielding an order of magnitude improvement in the tolerance of the analysis routine with regards to the single-frame active emitter density. Multiple-emitter fitting enables the overall performance of single-molecule super-resolution to be improved in one or more of several metrics that result in higher single-frame density of localized active emitters. For speed, the algorithm is implemented on Graphics Processing Unit (GPU) architecture, resulting in analysis times on the order of minutes. We show the performance of multiple emitter fitting as a function of the single-frame active emitter density. We describe the details of the algorithm that allow robust fitting, the details of the GPU implementation, and the other imaging processing steps required for the analysis of data sets. (C) 2011 Optical Society of America

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据