Within-day and between-day repeatability of measurements with an electronic nose in patients with COPD

期刊

JOURNAL OF BREATH RESEARCH
卷 7, 期 1, 页码 -

出版社

IOP PUBLISHING LTD
DOI: 10.1088/1752-7155/7/1/017103

关键词

-

资金

  1. PRIN
  2. Catholic University of the Sacred Heart, Fondi di Ateneo

向作者/读者索取更多资源

Electronic noses (e-noses), artificial sensor systems generally consisting of chemical sensor arrays for the detection of volatile compound profiles, have potential applications in respiratory medicine. We assessed within-day and between-day repeatability of an e-nose made from 32 sensors in patients with stable chronic obstructive pulmonary disease (COPD). We also compared between-day repeatability of an e-nose, fraction of exhaled nitric oxide (FENO) and pulmonary function testing. Within-day and between-day repeatability for the e-nose was assessed in two breath samples collected 30 min and seven days apart, respectively. Repeatability was expressed as an intraclass correlation coefficient (ICC). All sensors had ICC above 0.5, a value that is considered acceptable for repeatability. Regarding within-day repeatability, ICC ranged from 0.75 to 0.84 (mean = 0.80 +/- 0.004). Sensors 6 and 19 were the most reproducible sensors (both, ICC = 0.84). Regarding between-day repeatability, ICC ranged from 0.57 to 0.76 (mean = 0.68 +/- 0.01). Sensor 19 was the most reproducible sensor (ICC = 0.76). Within-day e-nose repeatability was greater than between-day repeatability (P < 0.0001). Between-day repeatability of FENO (ICC = 0.91) and spirometry (ICC range = 0.94-0.98) was greater than that of e-nose (mean ICC = 0.68). In patients with stable COPD, the e-nose used in this study has acceptable within-day and between-day repeatability which varies between different sensors.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据