Assessing the effect of CT slice interval on unidimensional, bidimensional and volumetric measurements of solid tumours

期刊

CANCER IMAGING
卷 12, 期 3, 页码 497-505

出版社

E-MED
DOI: 10.1102/1470-7330.2012.0046

关键词

Computed tomography (CT); response assessment; volume; measurement variability; slice interval

资金

  1. National Cancer Institute (NCI) [R01CA125143]
  2. AstraZeneca

向作者/读者索取更多资源

Objectives: To study the magnitude of differences in tumour unidimensional (1D), bidimensional (2D) and volumetric (VOL) measurements determined from computed tomography (CT) images reconstructed at 5, 2.5 and 1.25 mm slice intervals. Materials and Methods: A total of 118 lesions in lung, liver and lymph nodes were selected from 30 patients enrolled in early phase clinical trials. Each CT scan was reconstructed at 5, 2.5 and 1.25 mm slice intervals during the image acquisition. Lesions were semi-automatically segmented on each interval image series and supervised by a radiologist. 1D, 2D and VOL were computed based on the final segmentation results. Average measurement differences across different slice intervals were obtained using linear mixed-effects analysis of variance models. Results: Lesion diameters ranged from 6.1 to 80.1 mm (median 18.4 mm). The largest difference was seen between 1.25 and 5 mm (mean difference of 7.6% for 1D [P < 0.0001], 13.1% for 2D [P < 0.0001], -5.7% for VOL [P = 0.0001]). Mean differences between 1.25 and 2.5 mm were all within +/- 3.5% (within +/- 6% confidence interval). For VOL, there was a larger average difference between measurements on different slice intervals for the smaller lesions (<= 10 mm) compared with the larger lesions. Conclusions: Different slice intervals may give different 1D, 2D and VOL measurements. In clinical practice, it would be prudent to use the same slice interval for consecutive measurements.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据