4.6 Article

Noninvasive LV Pressure Estimation Using Subharmonic Emissions From Microbubbles

期刊

JACC-CARDIOVASCULAR IMAGING
卷 5, 期 1, 页码 87-92

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.jcmg.2011.08.017

关键词

contrast echocardiography; noninvasive pressure estimation; subharmonic microbubble signals

资金

  1. American Heart Association, Dallas, Texas [0655441U]
  2. National Institutes of Health, Bethesda, Maryland [R21 HL081892, RC1 DK087365]
  3. US Army Medical Research & Material Command, Fort Detrick, Maryland [W81XWH-08-1-0503]

向作者/读者索取更多资源

To develop a new noninvasive approach to quantify left ventricular (LV) pressures using subharmonic emissions from microbubbles, an ultrasound scanner was used in pulse inversion grayscale mode; unprocessed radiofrequency data were obtained with pulsed wave Doppler from the aorta and/or LV during Sonazoid infusion. Subharmonic data (in dB) were extracted and processed. Calibration factor(mmHg/dB) from the aortic pressure was used to estimate LV pressures. Errors ranged from 0.19 to 2.50mmHg when estimating pressures using the aortic calibration factor, and were higher (0.64 to 8.98 mmHg) using a mean aortic calibration factor. Subharmonic emissions from ultrasound contrast agents have the potential to noninvasively monitor LV pressures. (J Am Coll Cardiol Img 2012; 5: 87-92) (C) 2012 by the American College of Cardiology Foundation