4.2 Article

Checking and Improving Calibration of Raman Spectra using Chemometric Approaches

出版社

OLDENBOURG VERLAG
DOI: 10.1524/zpch.2011.0077

关键词

Raman Spectroscopy; Wavenumber Calibration; Intensity Calibration; Chemometrics

资金

  1. German Aerospace Centre [DLR 50OW0502]
  2. Federal Ministry of Education and Research, Germany (BMBF) [13N9205, FKZ13N9364]

向作者/读者索取更多资源

A wavenumber and intensity calibration procedure of Raman spectra by using chemometric techniques is presented. This approach allows the fine tuning of calibration parameters and routines with the final goal to eliminate setup dependent differences within experimentally recorded Raman spectra. This seems to be necessary since more and more Raman databases are needed for different analytical tasks, like identification of minerals or bacteria. Minimizing the impact of the applied experimental Raman setup on the reference (database stored) Raman spectra allows the databases to be enlarged very easily by feeding the database with Raman spectra recorded with different setups. Furthermore the chemometric analysis performance increases due to the larger number and better quality of reference spectra.