4.0 Article

Solid-phase preconcentration and determination of mercury(II) using activated carbon in drinking water by X-ray fluorescence spectrometry

期刊

X-RAY SPECTROMETRY
卷 42, 期 2, 页码 100-104

出版社

WILEY
DOI: 10.1002/xrs.2440

关键词

-

资金

  1. Instituto de Quimica de San Luis (INQUISAL)
  2. Universidad Nacional de San Luis
  3. Consejo Nacional de Investigaciones Cientificas y Tecnicas (CONICET)
  4. Agencia Nacional de Promocion Cientifica y Tecnologica (FONCYT) (PICT-BID)

向作者/读者索取更多资源

A novel and simple method was developed for the preconcentration and determination of mercury(II) from hydrochloric acid solutions through their adsorption on Aliquat 336 (tri-octylmethylammonium chloride)-activated carbon. The determinations were made directly on the solid by X-ray fluorescence spectrometry, which had the advantage of eliminating the step of elution of the mercury retained. This preconcentration system enabled the determination of Hg(II) in drinking water samples at mu gl1 levels. A 1700-fold enrichment factor was obtained. The described methodology showed excellent reproducibility, accuracy, and detection limits improvement by eliminating the step of elution of the analyte, replacing those methods based on cold vapor generation, reducing reagent consumption, and handling of samples. Copyright (c) 2013 John Wiley & Sons, Ltd.

作者