4.7 Article

Wear of materials used for artificial joints in total hip replacements

期刊

WEAR
卷 265, 期 1-2, 页码 143-149

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.wear.2007.09.010

关键词

total hip replacements; wear mechanisms; wear debris; corrosion potential; friction coefficient

向作者/读者索取更多资源

Two tribological pairs used for artificial joints were studied. Wear tests were performed in a pin-on-disc type machine modified to allow lubrication with bovine serum. The pins were made of either AISI F138 stainless steel or ASTM F136 titanium alloy, while the discs were manufactured with ultra-high molecular weight polyethylene (UHMWPE). The tested pairs were ASTM F138 stainless steel-UHMWPE and ASTM F136 titanium alloy-UHMWPE. The sliding velocity was fixed to 0.58 m/s for all the tests and the variation of friction force was registered as a function of time and normal load. Also, an electrochemical setup allowed monitoring the variation of the corrosion potential between the metallic pins and the bovine serum. The lower values of friction coefficient were measured when a thin film of polymer was transferred to the metallic surface, which led to a smooth interface and avoided debris generation. The viscoelastic behavior of UHMWPE was responsible for the increase in friction coefficient after a testing period, which depended on the normal load and the specific pair tested. (C) 2007 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据