4.5 Article

Potentiometric polymeric membrane electrodes for mercury detection using calixarene ionophores

期刊

WATER SCIENCE AND TECHNOLOGY
卷 61, 期 3, 页码 693-704

出版社

IWA PUBLISHING
DOI: 10.2166/wst.2010.860

关键词

Hg2+-ions; ion selective electrodes; potentiometry; PVC; p-tert-butyl-calix[4]arenethioether derivative ionophore

资金

  1. University Grant Commission (UGC)

向作者/读者索取更多资源

It is here established that potentiometric polymeric membrane electrodes based on electrically neutral ionophores are a useful analytical tool for the detection of heavy metal ions from environmental and industrial waste water. PVC based membrane containing p-tert-butyl-calix[4]arenethioether derivative as active material along with sodiumtetraphenylborate (NaTPB) as solvent mediator and dibutylphthalate as a plasticizer in the ratio 45:9:460:310 (w/w%) (I:NaTPB:DBP:PVC) exhibits good properties with a Nernstian response of 29.50 +/- 1.0 mV per decade of activity and a working concentration range of 7.2 x 10(-8)-1.0 x 10(-1) M. The electrode gave more stable potential readings when used around pH 2.5-6.8 and exhibits fast response time of 14 s. The sensors were found to work satisfactorily in partially non-aqueous media up to 40% (v/v) content of acetone, methanol or ethanol and could be used over a period of 7-9 months. Excellent selectivity for Hg2+ ions is indicated by match potential method and fixed interference method. The sensors could be used successfully in the estimation of mercury in different sample.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据