Comparison of chemistry analytes between 2 portable, commercially available analyzers and a conventional laboratory analyzer in reptiles

期刊

VETERINARY CLINICAL PATHOLOGY
卷 39, 期 4, 页码 474-479

出版社

WILEY
DOI: 10.1111/j.1939-165X.2010.00258.x

关键词

Clinical chemistry; i-STAT; method comparison; point-of-care; reptile; VetScan

资金

  1. Department of Small Animal Clinical Sciences, College of Veterinary Medicine, University of Tennessee

向作者/读者索取更多资源

Background Advantages of handheld and small bench-top biochemical analyzers include requirements for smaller sample volume and practicality for use in the field or in practices, but little has been published on the performance of these instruments compared with standard reference methods in analysis of reptilian blood. Objective The aim of this study was to compare reptilian blood biochemical values obtained using the Abaxis VetScan Classic bench-top analyzer and a Heska i-STAT handheld analyzer with values obtained using a Roche Hitachi 911 chemical analyzer. Methods Reptiles, including 14 bearded dragons (Pogona vitticeps), 4 blue-tongued skinks (Tiliqua gigas), 8 Burmese star tortoises (Geochelone platynota), 10 Indian star tortoises (Geochelone elegans), 5 red-tailed boas (Boa constrictor), and 5 Northern pine snakes (Pituophis melanoleucus melanoleucus), were manually restrained, and a single blood sample was obtained and divided for analysis. Results for concentrations of albumin, bile acids, calcium, glucose, phosphates, potassium, sodium, total protein, and uric acid and activities of aspartate aminotransferase and creatine kinase obtained from the VetScan Classic and Hitachi 911 were compared. Results for concentrations of chloride, glucose, potassium, and sodium obtained from the i-STAT and Hitachi 911 were compared. Results Compared with results from the Hitachi 911, those from the VetScan Classic and i-STAT had variable correlations, and constant or proportional bias was found for many analytes. Bile acid data could not be evaluated because results for 44 of 45 samples fell below the lower linearity limit of the VetScan Classic. Conclusions Although the 2 portable instruments might provide measurements with clinical utility, there were significant differences compared with the reference analyzer, and development of analyzer-specific reference intervals is recommended.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据