4.5 Article

MICROBUBBLE TYPE AND DISTRIBUTION DEPENDENCE OF FOCUSED ULTRASOUND-INDUCED BLOOD-BRAIN BARRIER OPENING

期刊

ULTRASOUND IN MEDICINE AND BIOLOGY
卷 40, 期 1, 页码 130-137

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.ultrasmedbio.2013.09.015

关键词

Blood-brain barrier opening; Focused ultrasound; Microbubble type; Microbubble distribution

资金

  1. National Institutes of Health [R01 EB009041, R01 AG038961]
  2. Kinetics Foundation
  3. NATIONAL INSTITUTE OF BIOMEDICAL IMAGING AND BIOENGINEERING [R01EB009041] Funding Source: NIH RePORTER
  4. NATIONAL INSTITUTE ON AGING [R01AG038961] Funding Source: NIH RePORTER

向作者/读者索取更多资源

Focused ultrasound, in the presence of microbubbles, has been used non-invasively to induce reversible blood-brain barrier (BBB) opening in both rodents and non-human primates. This study was aimed at identifying the dependence of BBB opening properties on polydisperse microbubble (all clinically approved microbubbles are polydisperse) type and distribution by using a clinically approved ultrasound contrast agent (Definity microbubbles) and in-house prepared polydisperse (IHP) microbubbles in mice. A total of 18 C57 BL/6 mice (n = 3) were used in this study, and each mouse was injected with either Definity or IHP microbubbles via the tail vein. The concentration and size distribution of activated Definity and IHP microbubbles were measured, and the microbubbles were diluted to 6 x 10(8)/mL before injection. Immediately after microbubble administration, mice were subjected to focused ultrasound with the following parameters: frequency = 1.5 MHz, pulse repetition frequency = 10 Hz, 1000 cycles, in situ peak rarefactional acoustic pressures = 0.3, 0.45 and 0.6 MPa for a sonication duration of 60 s. Contrast-enhanced magnetic resonance imaging was used to confirm BBB opening and allowed for image-based analysis. Permeability of the treated region and volume of BBB opening did not significantly differ between the two types of microbubbles (p > 0.05) at peak rarefractional acoustic pressures of 0.45 and 0.6 MPa, whereas IHP microbubbles had significantly higher permeability and opening volume (p < 0.05) at the relatively lower pressure of 0.3 MPa. The results from this study indicate that microbubble type and distribution could have significant effects on focused ultrasound-induced BBB opening at lower pressures, but less important effects at higher pressures, possibly because of the stable cavitation that governs the former. This difference may have become less significant at higher pressures, where inertial cavitation typically occurs. (C) 2014 World Federation for Ultrasound in Medicine & Biology.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据