INTER- AND INTRA-OPERATOR RELIABILITY AND REPEATABILITY OF SHEAR WAVE ELASTOGRAPHY IN THE LIVER: A STUDY IN HEALTHY VOLUNTEERS

期刊

ULTRASOUND IN MEDICINE AND BIOLOGY
卷 39, 期 6, 页码 950-955

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.ultrasmedbio.2012.12.011

关键词

Shear wave elastography; Reproducibility; Reliability; Repeatability; Ultrasound; Liver imaging; Intra-class correlation coefficient

向作者/读者索取更多资源

This study assessed the reproducibility of shear wave elastography (SWE) in the liver of healthy volunteers. Intra-and inter-operator reliability and repeatability were quantified in three different liver segments in a sample of 15 subjects, scanned during four independent sessions (two scans on day 1, two scans 1 wk later) by two operators. A total of 1440 measurements were made. Reproducibility was assessed using the intra-class correlation coefficient (ICC) and a repeated measures analysis of variance. The shear wave speed was measured and used to estimate Young's modulus using the Supersonics Imagine Aixplorer. The median Young's modulus measured through the inter-costal space was 5.55 +/- 0.74 kPa. The intra-operator reliability was better for same-day evaluations (ICC = 0.91) than the inter-operator reliability (ICC = 0.78). Intra-observer agreement decreased when scans were repeated on a different day. Inter-session repeatability was between 3.3% and 9.9% for intra-day repeated scans, compared with to 6.5%-12% for inter-day repeated scans. No significant difference was observed in subjects with a body mass index greater or less than 25 kg/m(2). (E-mail: hudsonjm@gmail.com) (C) 2013 World Federation for Ultrasound in Medicine & Biology.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据