ULTRASOUND IN MEDICINE AND BIOLOGY
卷 38, 期 5, 页码 745-752出版社
ELSEVIER SCIENCE INC
DOI: 10.1016/j.ultrasmedbio.2012.01.022
关键词
LIPUS; Ultrasound; Intensity; Fracture healing
资金
- Clinical Therapies R&D Department, Smith and Nephew Inc
Low-intensity pulsed ultrasound (LIPUS) with I-SATA = 30 mW/cm(2) has been proven in facilitating fracture healing, which the spatial average intensity over the on period (I-SATP) equals 150 mW/cm(2). As active ultrasound wave is only delivered during the on period, we postulate 150 mW/cm(2) is responsible for the beneficial effect of LIPUS. In this study, we compare the biologic effects of 30 mW/cm(2) and 150 mW/cm(2). We propose I-SATA = 150 mW/cm(2) could further enhance fracture healing process. Closed femoral fractured Sprague-Dawley rats were randomized into control, LIPUS-30 (30 mW/cm(2)) and LIPUS-150 (150 mW/cm(2)) groups. Weekly radiographs and endpoint microCT, histomorphometry, and biomechanical tests were performed. The results show that LIPUS-30 had significantly higher low-density bone volume fraction and woven bone percentage than that of control and LIPUS-150 in microCT and histologic measurements, respectively. Mechanically, failure torque of LIPUS-30 was significantly higher than control and LIPUS-150 at week 6. In conclusion, LIPUS at I-SATA = 150 mW/cm(2) did not further enhance fracture healing. (E-mail: louis@ort.cuhk.edu.hk) (C) 2012 World Federation for Ultrasound in Medicine & Biology.