4.2 Article

The Effect of Lumbar Lordosis on Screw Loosening in Dynesys Dynamic Stabilization: Four-Year Follow-Up with Computed Tomography

期刊

BIOMED RESEARCH INTERNATIONAL
卷 2015, 期 -, 页码 -

出版社

HINDAWI LTD
DOI: 10.1155/2015/152435

关键词

-

向作者/读者索取更多资源

Introduction. This study aimed to evaluate the effects of Dynesys dynamic stabilization (DDS) on clinical and radiographic outcomes, including spinal pelvic alignment. Method. Consecutive patients who underwent 1- or 2-level DDS for lumbar spondylosis, mild degenerative spondylolisthesis, or degenerative disc disease were included. Clinical outcomes were evaluated by Visual Analogue Scale for back and leg pain, Oswestry Disability Index, and the Japanese Orthopedic Association scores. Radiographic outcomes were assessed by radiographs and computed tomography. Pelvic incidence and lumbar lordosis (LL) were also compared. Results. In 206 patients with an average follow-up of 51.1 +/- 20.8 months, there were 87 screws (8.2%) in 42 patients (20.4%) that were loose. All clinical outcomes improved at each time point after operation. Patients with loosened screws were 45 years older. Furthermore, there was a higher risk of screw loosening in DDS involving S1, and these patients were more likely to have loosened screws if the LL failed to increase after the operation. Conclusions. The DDS screw loosening rate was overall 8.2% per screw and 20.4% per patient at more than 4 years of follow-up. Older patients, S1 involvement, and those patients who failed to gain LL postoperatively were at higher risk of screw loosening.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据