4.7 Review

Calibration and measurement issues for therapeutic ultrasound

期刊

ULTRASONICS
卷 48, 期 4, 页码 234-252

出版社

ELSEVIER
DOI: 10.1016/j.ultras.2007.10.010

关键词

ultrasound; calibration; measurement; physiotherapy; HIFU; HITU; focused ultrasound

资金

  1. National Measurement System Directorate of the UK Department of Trade and Industry

向作者/读者索取更多资源

This review paper examines some of the issues relating to calibration and measurement of therapeutic medical ultrasonic equipment (MUE). This is not intended to be an all-encompassing review of all aspects of characterising therapeutic ultrasound. Instead it concentrates on issues related to the acoustic output of two applications: physiotherapy and high intensity focused ultrasound surgery (HIFUS or HIFU; also referred to as high intensity therapeutic ultrasound, HITU). Physiotherapy has a well-established standards infrastructure for calibration: the requirements are small in number and well-defined. The issue for physiotherapy is not so much 'How to calibrate?' but rather, 'How to ensure that equipment IS calibrated?' The situation in the much newer area of HIFU is very different: the first steps towards writing standards are just starting and even the very basic questions of what to measure and with what type of sensor are open for debate. Readers whose main interest is in other ultrasound therapies will find ideas of relevance to their own speciality. Crown Copyright (C) 2007 Published by Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据