4.4 Article Proceedings Paper

Blind deconvolution of time-of-flight mass spectra from atom probe tomography

期刊

ULTRAMICROSCOPY
卷 132, 期 -, 页码 60-64

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.ultramic.2013.03.015

关键词

Atom probe tomography; Time-of-flight mass spectrometry; Blind deconvolution

资金

  1. Vinnova VINN Excellence Center on Functional Nanoscale Materials

向作者/读者索取更多资源

A major source of uncertainty in compositional measurements in atom probe tomography stems from the uncertainties of assigning peaks or parts of peaks in the mass spectrum to their correct identities. In particular, peak overlap is a limiting factor, whereas an ideal mass spectrum would have peaks at their correct positions with zero broadening. Here, we report a method to deconvolute the experimental mass spectrum into such an ideal spectrum and a system function describing the peak broadening introduced by the held evaporation and detection of each ion. By making the assumption of a linear and time-invariant behavior, a system of equations is derived that describes the peak shape and peak intensities. The model is fitted to the observed spectrum by minimizing the squared residuals, regularized by the maximum entropy method. For synthetic data perfectly obeying the assumptions, the method recovered peak intensities to within +/- 0.33 at%. The application of this model to experimental APT data is exemplified with Fe-Cr data. Knowledge of the peak shape opens up several new possibilities, not just for better overall compositional determination, but, e.g., for the estimation of errors of ranging due to peak overlap or peak separation constrained by isotope abundances. (C) 2013 Elsevier By. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据