4.5 Article

Rapid analysis of bacterial contamination of tap water using isothermal calorimetry

期刊

THERMOCHIMICA ACTA
卷 543, 期 -, 页码 273-280

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.tca.2012.06.002

关键词

Bacterial contamination; Calorimetry; Drinking water analysis; E. coli; P. putida

向作者/读者索取更多资源

Highly sensitive calorimetry was evaluated for its potential for the monitoring of the bacterial contamination of drinking water. For that purpose, water samples were added to different media and bacterial growth was followed microcalorimetrically. Drinking water samples were either tested untreated or after artificial contamination with selected bacterial strains. Two bacteria, two growth media and two growth conditions (i.e. aerobic and anaerobic) were applied. Even minor initial bacterial contamination (1-100 cells mL(-1)) gave rise to calorimetric signals after 5-17 h growth. Calorimetry was thus faster than detection of bacterial contamination by plating on agar and inspection of colony formation, which was performed for comparison. It was further demonstrated theoretically that calorimetric detection is superior to colony detection. The heat production curves were characteristic for the strain, the medium and the growth conditions. It is hypothesized that a further refinement of the microcalorimetric method via the application of sets of specific media in combination with selective growth conditions should allow delimiting the identity of the contaminant. (C) 2012 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据