THERAPEUTIC DRUG MONITORING
卷 31, 期 2, 页码 198-204出版社
LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/FTD.0b013e31819c6a37
关键词
tacrolimus; FK 506; immunoassay; LC-MSMS; method evaluation; therapeutic drug monitoring
资金
- Abbott Laboratories
The objective of this study was to evaluate the analytical performance of the Abbott ARCHITECT Tacrolimus immunoassay. Proficiency panels and specimens from a population of organ transplant recipients were analyzed in 6 clinical laboratories ill Europe and the United States, and the results were compared with other methods. The ARCHITECT assay requires a whole blood specimen pretreatment step with methanol/zinc sulfate to precipitate protein and extract the drug, followed by a 30-minute immunoassay using anti-tacrolimus antibody-coated paramagnetic microparticles and an acridinium-tacrolimus tracer. The assay was free from hematocrit interference in the range 25%-55% and from interference by extremes of cholesterol, triglycerides, bilirubin, total protein, and uric acid. The total percent of coefficient of variations of the assay were 4.9%-7.6% at 3 ng/mL, 2.9%-4.6% at 8.6 ng/mL, and 3.1%-8.2% at 15.5 ng/mL. Limit of detection was <= 0.5 ng/mL and limit of quantification (LOQ) ranged from 0.69 to 1.07 ng/mL across the 6 sites (based on the upper 95% confidence interval concentrations). The 2007 European Consensus Conference oil Tacrolimus Optimization recommended the use of assay methods with an LOQ around I ng/mL, based upon the need to measure trough tacrolimus blood concentrations precisely down to 3 ng/mL during low-dose tacrolimus regimens. Tacrolimus International Proficiency Testing Scheme samples were measured by the ARCHITECT immunoassay at 5 sites and showed an average bias of -0.28 to +0.85 ng/mL versus IMx Tacrolimus 11 immunoassay historical values and -0.21 to +0.68 ng/mL versus liquid chromatography/tandem mass spectrometry (LC-MSMS) Tacrolimus historical values. Method comparison studies were performed with the ARCHITECT Tacrolimus immunoassay on patient specimens with the following results: ARCHITECT Tacrolimus assay versus the Abbott IMx Tacrolimus 11 immunoassay (4 sites) yielded average biases between -0.94 and +0.26 ng/mL; ARCHITECT assay versus the Dade Dimension Tacrolimus immunoassay (2 sites) yielded average biases of -0.46 and +0.11 ng/mL; and ARCHITECT assay versus LC-MSMS methods at 2 sites yielded average biases of +0.51 and +1.63 ng/mL. Spearman correlation coefficients were >= 0.90 on all method comparisons. The ARCHITECT Tacrolimus assay is a semiautomated, robust, and highly sensitive immunoassay, representing an alternative approach for laboratories not equipped with LC-MSMS, and meets the I ng/mL recommendation of LOQ by the European Consensus Conference on Tacrolimus Optimization.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据