期刊
TALANTA
卷 129, 期 -, 页码 523-528出版社
ELSEVIER
DOI: 10.1016/j.talanta.2014.06.021
关键词
Graphene-platinum hybrid nanoparticle; GCPE; Genosensor; E. coli
资金
- Mugla Sitki Kocman University BAP Project [12/115]
Glassy carbon paste electrode (GCPE) was modified with graphene platinum hybrid nanoparticle (Gr-Pt hybrid NP) and used as a transducer for label and indicator free electrochemical genosensor. 22 mer oligonucleotides representing Escherichia coli bacteria were used as a model case. As far as it is known, this study is the first study where Gr-Pt hybrid NP was incorporated into GCPE and used for genosensor transducer. The extent of hybridization was determined by using differential pulse voltammetric signals of guanin oxidation. After the optimization of experimental parameters, analytical characteristics were investigated. The linear range was found between 1.5 x 10(-7) and 2.25 x 10(-6) M with the equation of y= 1.6566x-2.6161 and R-2 of 0.9959. RSD and LOD were calculated as 4.2% (n=6) and 1.12 x 10(-9) M respectively. (C) 2014 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐