期刊
TALANTA
卷 88, 期 -, 页码 468-476出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.talanta.2011.11.016
关键词
Horseradish peroxidase; Square wave voltammetry; Peroxidase cosubstrates; o-Phenilendiamine; Response surface methodology
资金
- Universidad Nacional del Litoral Santa Fe, Argentina [8/41]
The electrochemical detection for horseradish peroxidase-cosubstrate-H2O2 systems was optimized. o-Phenilendiamine, phenol, hydroquinone, pyrocatechol, p-chlorophenol, p-aminophenol and 3,3'-5,5'-tetramethylbenzidine were evaluated as cosubstrates of horseradish peroxidase (HRP) enzyme. Therefore, the reaction time, the addition sequence of the substrates, the cosubstrate:H2O2 ratio and the electrochemical techniques were elected by one-factor optimization assays while the buffer pH, the enzymatic activity and cosubstrate and H2O2 concentrations for each system were selected simultaneously by response surface methodology. Then, the calibration curves for seven horseradish peroxidase-cosubstrate-H2O2 systems were built and the analytic parameters were analyzed. o-Phenilendiamine was selected as the best cosubstrate for the HRP enzyme. For this system the reaction time of 60 s. the phosphate buffer pH 6.0, and the concentrations of 2.5 x 10(-4) mol L-1 o-phenilendiamine and of 1.25 x 10(-4) mol L-1 H2O2 were chosen as the optimal conditions. In these conditions, the calibration curve of horseradish peroxidase by square wave voltammetry showed a linearity range from 9.5 x 10(-11) to 1.9 x 10(-8) mol L-1 and the limit of detection of 3.8 x 10(-11) mol L-1 with RSD% of 0.03% (n = 3). (C) 2011 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据