期刊
TALANTA
卷 86, 期 -, 页码 400-407出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.talanta.2011.09.034
关键词
Gold nanoparticles; Self-assembly; Amperometry; Bilirubin; Blood serum
资金
- Academic Research Funding Tier 1 [RG65/08]
- National Medical Research Council of Singapore [NMRC/NIG/00602009]
We describe the development of a simple and highly sensitive electrochemical (amperometric) sensing of bilirubin based on bilirubin oxidase (BOx) incorporated into the gold nanoparticles (AuNPs). This nanoelectrode platform with self-assembled enzyme is highly sensitive toward the electrochemical oxidation of bilirubin and increased the bilirubin concentration linearly from to 5000 mu M with a correlation coefficient of 0.9960, and an apparent Michaelis constant (K-M,K-app) of 44 +/- 0.4 mu M. Using an amperometric method, the detection limit for bilirubin at the enzyme-modified electrode was 1.4 nM (signal-to-noise ratio = 3). The modified electrode retained a stable response for 2 days while losing only ca. 3.4% of its initial sensitivity during a 10 days storage period in 0.2 M phosphate buffer solution (pH = 8.4) at <= 4 degrees C. The practical application of the modified electrode was demonstrated by measuring the concentration of bilirubin in blood serum sample. (C) 2011 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据