4.7 Article

Chemiluminescence detection of label-free C-reactive protein based on catalytic activity of gold nanoparticles

期刊

TALANTA
卷 84, 期 3, 页码 752-758

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.talanta.2011.02.001

关键词

C-reactive protein; Gold nanoparticles (GNPs); Chemiluminescence detection

资金

  1. Korea Ministry of Science and Technology [M1053608003-05N3608-00310]

向作者/读者索取更多资源

A novel, quantitative analytical method for measuring C-reactive protein (CRP) levels in human serum has been developed based on the catalytic activity of gold nanoparticles (GNPs) and luminol-H2O2 chemiluminescence (CL). The CL intensity in the presence of CRP and its ligand, O-phosphorylethanolamine (PEA), was greatly enhanced due to the aggregation of GNPs after the addition of 0.5 M NaCl. Any pretreatment steps, such as covalent functionalization of GNPs, addition of antibodies, or labeling of CRP, were not needed for CL detection. The CL enhancement was linearly proportional to CRP concentration in the range of 1.88 fM to 1.925 pM. The detection limit of CRP in serum samples was estimated to be as low as 1.88 fM. The detection sensitivity was increased more than 164 times of magnitude over that of the conventional, enzyme-linked immunosorbent assay (ELISA) method. This proposed GNP-based CL detection method offers the advantages of simplicity, rapidity, and sensitivity. (C) 2011 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据