A comparison of the ion chemistry for mono-substituted toluenes and anilines by three methods of atmospheric pressure ionization with ion mobility spectrometry

期刊

TALANTA
卷 78, 期 4-5, 页码 1464-1475

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.talanta.2009.02.043

关键词

Ion mobility spectrometry; Atmospheric pressure ionization; Mass spectrometry; Isomeric compounds

资金

  1. Deutsche Forschungsgemeinschaft (DFG)

向作者/读者索取更多资源

Ion mobility spectra for a series of mono-substituted toluenes and a series of mono-substituted anilines were obtained using three different methods of atmospheric pressure ionization including photoionization, chemical ionization from a (63)Ni source, and chemical ionization from a corona discharge source. The product ion peak intensities were measured as functions of analyte concentration at 323 K in a purified air atmosphere. Two, and sometimes three, product ion peaks were observed in spectra from chemical ionization with the (63)Ni source and it is suggested that the major peak, due to the protonated molecule, arose in both series by proton transfer from H(3)O(+)(H(2)O)(n). The second peak with diminished intensity and longer drift time than the protonated molecule can be seen with the toluenes and Was understood to be the NO(+) adduct, formed from the reactant ion NO(+)(H(2)O)(n). Electron transfer from the anilines to the latter ion yields the molecular ions, identified by having the same reduced mobility coefficients as the molecular ions produced by photoionization. The structure of these product ions was determined by investigations using the coupling of ion mobility spectrometry with atmospheric pressure photoionization and mass spectrometry (APPI-IMS-MS). The relative abundances of both the NO(+) adducts with the toluenes and the molecular ions with the anilines are enhanced with a corona discharge source where relatively more NO(+)(H(2)O)(n) is produced than in a (63)Ni source. Ab initio calculations show that only the protonated anilines of all the product ions are significantly hydrated with 1 ppm(v) of moisture in the supporting atmosphere of the ion mobility spectrometer. (C) 2009 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据