4.7 Article

The effect of signal suppression and mobile phase composition on the simultaneous analysis of multiple classes of acidic/neutral pharmaceuticals and personal care products in surface water by solid-phase extraction and ultra performance liquid chromatography-negative electrospray tandem mass spectrometry

期刊

TALANTA
卷 74, 期 5, 页码 1299-1312

出版社

ELSEVIER
DOI: 10.1016/j.talanta.2007.08.037

关键词

pharmaceuticals and personal care products; ultra performance liquid chromatograpby-tandem mass spectrometry; solid-phase extraction; multi-residue method; ion suppression; mobile-phase additives

向作者/读者索取更多资源

Anew multi-residue method for the determination of 25 acidic/neutral pharmaceuticals (antibiotics, anti-inflammatory/analgesics, lipid regulating agents, diuretics, triazides, H2-receptor antagonists, cardiac glicozides and angiotensin II antagonists) and personal care products (sunscreen agents and preservatives) in surface water with the usage of a new technique: ultra performance liquid chromatography-negative electrospray tandem mass spectrometry (lJPLC-MS/MS) was developed and validated. The novel UPLC system with 1.7 mu m particle-packed column allowed for good resolution of analytes with the application of low mobile phase flow rates (0.05 mL min(-1)) and short retention times (from 4.7 min to 13.3 min) delivering a fast and cost-effective multi-residue method. SPE with the usage of Oasis MCX strong cation-exchange mixed-mode polymeric sorbent was chosen for sample clean-up and concentration. The influence of mobile-phase composition, matrix assisted ion suppression and SPE recovery on the sensitivity of the method was identified and quantified. The instrumental limits of quantification varied from 0.2 mu g L-1 to 30 mu g L-1. The method limits of quantification were at low nanogram per litre levels and ranged from 0.3 ng L-1 to 30 ng L-1. The instrumental and method intra-day and inter-day repeatabilities were on average less than 5%. The method was successfully applied for the determination of PPCPs in River Taff. Thirteen compounds were determined in river water at levels ranging from a single to a few hundred nanograms per litre. Among them were ten pharmaceuticals (aspirin, salicylic acid, ketoprofen, naproxen, diclofenac, ibuprofen, mefenamic acid, furosemide, sulfasalazine and valsartan) and three personal care products (methyl- and ethylparaben and 4-benzophenone). (c) 2007 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据