Interlaboratory comparison study of serum total testoserone measurements performed by mass spectrometry methods

期刊

STEROIDS
卷 74, 期 6, 页码 498-503

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.steroids.2009.01.004

关键词

Testosterone; HPLC-MS/MS; Method comparison; Analytical variability

资金

  1. Solvay Pharmaceutical through the CDC Foundation
  2. National Center for Chronic Disease Prevention and Health Promotion

向作者/读者索取更多资源

Background: Though mass spectrometry (MS) assays are increasingly used for routine clinical measurements of serum total testosterone (TT), information about the variability of results is limited. This study assessed the variability of TT measurement results from routine MS assays. Methods: Twenty serum samples (12 females, 8 males) were analyzed on 2 days by seven high performance liquid chromatography (HPLC), and one gas chromatography (GC)-tandem mass spectrometry (HPLC-MS/MS, CC-MS/MS) assays. Two samples (male and female) were provided in five replicates to assess the within-run variability. Results were compared against those obtained at National Institute of Standards and Technology (NIST). The within- and between-laboratory variability was assessed for each sample. Comparisons to the NIST results were performed using bias plot and Deming regression analysis. Results: The overall coefficient of variation of the results obtained with MS assays was <15%CV at >1.53 nmol/L and <34%CV at 0.3 nmol/L The between-assay variability was the major contributor to the overall variability. The assay precision was the highest (<3%CV) with assays using liquid-liquid extraction for sample preparation or GC-MS/MS. The mean percent difference to the reference assay was 11%. The slopes of Deming regression analysis of the MS assays were between 0.903 and 1.138 (correlation coefficient: >0.996). TT concentrations for one assay were above the measurement range. Conclusions: The variability of TT measurement results among MS assays is substantially smaller than that reported for immunoassays. The type of sample preparation may affect assay precision. Standardizing assays can further reduce the variability of measurement results. Published by Elsevier Inc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据