期刊
STEROIDS
卷 74, 期 12, 页码 920-926出版社
ELSEVIER SCIENCE INC
DOI: 10.1016/j.steroids.2009.06.007
关键词
Testosterone; 5 alpha-Dihydrotestosterone; Liquid chromatography-electrospray ionization tandem mass spectrometry; Picolinyl derivatization; Prostatic disorder diagnosis
资金
- Ministry of Education, Culture, Sports, Science and Technology of Japan
We developed highly sensitive detection of testosterone (T) and 5 alpha-dihydrotestosterone (DHT) by liquid chromatography-electrospray ionization tandem mass spectrometry using high proton affinitive derivatization of 17 beta-hydroxyl group of T and DHT with picolinic acid, mobile phase consisting of MeCN-MeOH-H2O-formic acid and conventional octadecylsilica (ODS) column. Purification of the derivatives was carried out using solid-phase extraction with ODS cartridge. By this method, T and DHT were determined simultaneously with limits of quantification (LOQs) of 1 pg/0.2 ml in serum, and T and DHT with LOQs of 0.5 pg and 1 pg/3 mg in prostate tissue, respectively, under acceptable assay performance (intra-assay and inter-assay accuracy and precision). The present method provides reliable and reproducible results for quantification of T and DHT in small volumes of serum and prostate samples for diagnosis in prostatic disorders and male climacteric. (C) 2009 Elsevier Inc. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐