4.5 Article

Validation of a novel minimally invasive intervertebral disc pressure sensor utilizing in-fiber Bragg gratings in a porcine model -: An ex vivo study

期刊

SPINE
卷 33, 期 17, 页码 E589-E594

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/BRS.0b013e31817c55e2

关键词

spine; intervertebral disc; back pain; discography; nucleus pulposus; disc degeneration; pressure sensor

向作者/读者索取更多资源

Study Design. Nucleus pressure was measured within porcine intervertebral discs (IVDs) with a novel in-fiber Bragg grating (FBG) sensor (0.4 mm diameter) and a strain gauge (SG) sensor (2.45 mm). Objective. To validate the accuracy of a new FBG pressure sensor designed for minimally invasive measurements of nucleus pressure. Summary of Background Data. Although its clinical utility is controversial, it is possible that the predictive accuracy of discography can be improved with IVD pressure measurements. These measurements are typically obtained using needle-mounted SG sensors inserted into the nucleus. However, by virtue of their size, SG sensors alter disc mechanics, injure anulus fibers, and can potentially initiate or accelerate degenerative changes thereby limiting their utility particularly clinically. Methods. Six functional spinal units were loaded in compression from 0 N to 500 N and back to 0 N; nucleus pressure was measured using the FBG and SG sensors at various locations along anterior and anterolateral axes. Results. On average maximum IVD pressures measured using the FBG and SG sensors were within 9.39% of each other. However, differences between maximum measured pressures from the FBG and SG sensors were larger (22.2%) when the SG sensor interfered with vertebral endplates (P < 0.05). The insertion of the FBG sensor did not result in visible damage to the anulus, whereas insertion of the SG sensor resulted in large perforations in the anulus through which nucleus material was visible. Conclusion. The new FBG sensor is smaller and less invasive than any previously reported disc pressure sensor and gave results consistent with previous disc pressure studies and the SG sensor. There is significant potential to use this sensor during discography while avoiding the controversy associated with disc injury as a result of sensor insertion.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据