Common analyte internal standardization as a tool for correction for mass discrimination in multi-collector inductively coupled plasma-mass spectrometry

期刊

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.sab.2013.08.009

关键词

Boron; Antimony; Isotope ratio; Multi-collector ICP-MS; Mass discrimination correction

资金

  1. ERC Starting Grant ARCHGLASS [240750]
  2. Ghent University Special Research Fund (BOF UGent)
  3. FWO [G002111N]

向作者/读者索取更多资源

It is well known that to achieve accurate isotope ratio data using multi-collector inductively coupled plasma-mass spectrometry (MC-ICP-MS), mass discrimination needs to be adequately corrected for. In this work, the capabilities and limitations of common analyte internal standardization (CAIS) as a method for correction for mass discrimination were assessed for two target elements, one in the low mass region (boron) and another in the medium mass region (antimony). CAIS has already been used in the context of element determination and, more recently, also in isotope ratio measurements with quadrupole-based ICP-MS, but it has, to the best of the authors' knowledge, never been applied to isotope ratio determination via MC-ICP-MS so far. Results obtained relying on CAIS were compared with those obtained via more established correction models, i.e. external correction in a standard-sample bracketing approach (SSB) and internal correction based on both the Russell's law (original and empirical) and the revised Russell's law. To the best of the authors' knowledge, such comparison has not been done before, especially not for low mass elements. Also the robustness of these approaches with respect to concomitant matrix elements was assessed using synthetic solutions, containing various concentrations of the matrix elements Be, Cs, Be + Cs or Fe. Whereas for B, the isotope ratio result did not seem to be significantly affected by the matrix, thus suggesting that complete separation of B from a matrix might not be necessary, at least in the cases studied, the mass discrimination observed for Sb was influenced by the presence of Cs. The experiments carried out demonstrate that the CAIS technique can be successfully applied for mass bias correction in MC-ICP-MS, providing data of the same quality as the revised Russell's law, while being more transparent and accessible. While for B, all mass bias correction approaches tested provided similar data quality (external correction with SSB gives slightly better precision), for Sb, CAIS and the revised Russell's law provide better precision and accuracy. (C) 2013 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据