4.6 Article

Femtosecond laser ablation: Experimental study of the repetition rate influence on inductively coupled plasma mass spectrometry performance

期刊

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.sab.2007.11.035

关键词

LA-ICP-MS; femtosecond LA-ICP-MS; repetition rate effects

向作者/读者索取更多资源

This paper demonstrates the feasibility of performing bulk chemical analysis based on laser ablation for good lateral resolution with only nominal mass ablated per pulse. The influence of repetition rate (1-1000 Hz) and scan speed (1-200 mu m/s) using a low energy (30 mu J) and a small spot size (similar to 10 mu m) UV-femtosecond laser beam was evaluated for chemical analysis of silica glass samples, based on laser ablation sampling and inductively coupled plasma mass spectrometry (ICP-MS). Accuracy to approximately 14% and precision of 6% relative standard deviation (RSD) were measured. Published by Elsevier B.V.