4.7 Article

Speciation of arsenic (III) and arsenic (V) based on quenching of CdS quantum dots fluorescence using hybrid sequential injection-stopped flow injection gas-diffusion system

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.saa.2012.05.054

关键词

Hybrid SI-Fl; Gas diffusion; Arsenic speciation; Cadmium sulfide quantum dots; Mercaptoacetic acid

资金

  1. Higher Education Research Promotion and National Research University Project of Thailand
  2. Office of the Higher Education Commission
  3. Rajamangala University of Technology Isan (RMUTI)
  4. Center of Excellence for Innovation in Chemistry (PERCH-CIC), Commission on Higher Education, Ministry of Education

向作者/读者索取更多资源

A hybrid sequential injection-stopped flow injection system was developed for the speciation of arsenic based on the quenching of mercaptoacetic acid capped cadmium sulfide quantum clots (CdS-MAA QDs) fluorescence intensity. The analytical procedure involves the generation of arsine from As(III) by sodium borohydride in acetate buffer medium pH 6.0. The generated arsine (donor stream) diffuses across the PTFE membrane of the gas-diffusion unit into an acceptor stream and then interacts with CdS-MAA QDs. Total arsenic was determined after pre-reduction of As(V) to As(III) with 1% (m/v) mercaptoacetic acid. Concentration of As(V) in the sample solutions can be deduced from the difference of total arsenic and As(III). Optimization of the experimental conditions and instrumental parameters were investigated. Under optimal conditions, limits of detection were 20 mu g L-1 for As(III) and 40 mu g L-1 for As(V). Recoveries in the range 84-103% were obtained from sediment sample. (C) 2012 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据