A simple method has been developed for the preconcentration of mercury based on the adsorption of its diphenylthiocarbazone complex on a neutral alumina column. The influence of acidity, eluting agents, stability of the column, sample volume and interfering ions has been investigated in detail. The adsorbed complex could be eluted using environmentally benign polyethylene glycol (PEG 400) and the concentration of mercury was determined by visible spectrophotometry at a wavelength maximum of 520 nm. A detection limit of 4 mu g L-1 could be achieved and the developed procedure was successfully applied for the determination of mercury in spiked water samples and city waste incineration ash (CRM176). The preconcentration factor attainable for quantitative recovery (>95%) of mercury(II) was 100 for a 1000 mL sample volume. (C) 2007 Elsevier B.V. All rights reserved.