4.6 Article

A geometrical optimization of a magneto-rheological rotary brake in a prosthetic knee

期刊

SMART MATERIALS AND STRUCTURES
卷 19, 期 3, 页码 -

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0964-1726/19/3/035023

关键词

-

资金

  1. University of Iceland
  2. Ossur Inc

向作者/读者索取更多资源

Magneto-rheological (MR) fluids have been successfully introduced to prosthetic devices. One such device is a biomechanical prosthetic knee that uses MR fluids to actively control its rotary stiffness. The brake is rotational, utilizing the MR fluid in shear mode. In this study, the geometrical design of the MR brake is addressed. This includes the design of the magnetic circuit and the geometry of the fluid chamber. Mathematical models are presented that describe the rotary torque of the brake. A novel perfluorinated polyether (PFPE)-based MR fluid is introduced, whose properties are tailored for the prosthetic knee. On-state and off-state rheological measurements of the MR fluid are presented. The finite element method is used to evaluate the magnetic flux density in the MR fluid. The design is formulated as an optimization problem, aiming to maximize the braking torque. A parametric study is carried out for several design parameters. Subsequently, a multi-objective optimization problem is defined that considers three design objectives: the field-induced braking torque, the off-state rotary stiffness and the weight of the brake. Trade-offs between the three design objectives are investigated which provides a basis for informed design decisions on furthering the success of the MR prosthetic knee.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据