4.3 Article

Reliability of Repeated Measurements on Post-Burn Scars with Corneometer CM 825®

期刊

SKIN RESEARCH AND TECHNOLOGY
卷 21, 期 3, 页码 302-312

出版社

WILEY
DOI: 10.1111/srt.12193

关键词

burn scar; reliability; water content; water holding capacity; stratum corneum

向作者/读者索取更多资源

Background/purposeThe water content in burn scars, the parameter of stratum corneum water holding capacity, is an important feature in evaluation of biophysical properties of scars. Nevertheless, quantifiying this parameter is a challenge. In this study, the reliability of repeated water content measurements with Corneometer CM825((R)) on (burn) scars was investigated. MethodsIntra-observer reliability, inter-observer reliability and day-by-day variability were examined on 30 scars by means of intra-class correlation coefficient (ICC) and within-subject coefficient of variation (WSCV). Bland-Altman plots with 95% limits of agreement' were constructed. ResultsResults revealed excellent ICC values (ICCintra=0.985; ICCinter=0.984) with relatively low WSCV (WSCVintra=6.3%; WSCVinter=10.6%) for respectively intra- and inter-observer reliability. However, the Bland-Altman plot showed that more than 5% of differences were expected to exceed 4a.u., the limit of what has been defined as a clinically acceptable difference. Results for day-by-day variability showed good ICC value (ICCday-by-day=0.849) and higher WSCV (WSCVday-by-day=20.5%). ConclusionThe Corneometer CM825((R)) is an objective and sensitive instrument for water content measurements. On the basis of our results, we concluded that the instrument can be used in clinical trials, but only under very strict conditions with standardized test protocol, preferably in combination with the evaluation of other physiological parameters.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据