4.7 Article

Fabrication and characterization of a surface plasmon resonance based fiber optic urea sensor for biomedical applications

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 161, 期 1, 页码 434-438

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2011.10.056

关键词

Surface plasmon; Optical fiber; Sensor; Urea; Urease

资金

  1. Department of Science and Technology, India
  2. Council for Scientific and Industrial Research, India

向作者/读者索取更多资源

The fabrication and characterization of a surface plasmon resonance based fiber optic sensor for the detection of urea in liquid are reported. The probe is fabricated by coating three layers namely of silver, silicon and the enzyme, urease, over about 1 cm length of the unclad core of an optical fiber. The silicon layer protects silver from oxidation and enzyme from deactivation in addition to enhance the sensitivity of the sensor. The wavelength interrogation method is used to characterize the sensor. It is observed that the resonance wavelength decreases as the concentration of urea increases. The sensor operates in the urea concentration range of 0-160 mM, which is close to the physiological range of urea in blood and hence can be used in medical sciences. Further, the effects of concentration of urea on the sensitivity and detection accuracy of the sensor are studied. It is found that the sensitivity decreases with the increase in the concentration of urea whereas the detection accuracy is nearly independent of the urea concentration. (C) 2011 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据