Prothrombin time tests on a microfluidic disc analyzer

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 161, 期 1, 页码 1184-1190

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2011.11.025

关键词

Prothrombin time test; International normalized ratio; Microfluidic disc analyzer; Centrifugation; Mixing

资金

  1. National Science Council (NSC) [98-2221E-035-030]
  2. Feng Chia University
  3. Taichung Veterans General Hospital [TCVGH-FCU988205]

向作者/读者索取更多资源

Prothrombin time test is used to measure the time required for blood to coagulate and is essential for monitoring appropriate drug dosing in patients taking oral anti-coagulants. To reduce the costs of reagents and improve the stability of tests, a microfluidic disc analyzer (MDA) has been developed. This analyzer can process small volumes of liquid samples (less than 10 mu l total volume required) and performs effective and fast mixing (within one second) under centrifugation. During the mixing process, two reactants can also maintain a constant ratio of flowrate. One hundred and one clinical samples were tested by the MDA and the Sysmex CA1500 coagulation analyzer, which was frequently used as a reference machine, for comparison. The test results from these two instruments show good correlation and agreement. (c) 2011 Elsevier B.V. All rights reserved.